Safety DF-310E 107
11 Safety
CAUTION
Do not setup or operate the Oxygen Analyzer without a complete
understanding of the instructions in this manual. Do not connect
this Analyzer to a power source until all signal and plumbing
connections are made.
CAUTION
This analyzer must be operated in a manner consistent with its
intended use and as specified in this manual.
DANGER
Potentially hazardous AC voltages are present within this
instrument. Leave all servicing to qualified personnel. Disconnect
the AC power source when installing or removing: external
connections, the sensor, the electronics, or when charging or
draining electrolyte.
DANGER
The electrolyte is a caustic solution. Review the Material
Safety Data Sheet (MSDS) before handling the
electrolyte solution.
The sensor is shipped dry and must be charged with
electrolyte before it is operated.
CAUTION
Over-pressurizing the sensor can result in permanent damage to the
sensor. Limit the backpressure to the analyzer to ±1 psig.
Be sure the downstream isolation valve (if so equipped) is toggled
open before gas flow is started.
CAUTION
DO NOT SHIP THE ANALYZER WITH
ELECTROLYTE THOROUGHLY DRAIN AND
RINSE SENSOR BEFORE SHIPPING