PERFORMANCE
ACCURACY
Standard Resolution: Greater of ± 3% of reading (not to exceed 1% of
range for % Analyzers) or 0.5% of range.
High Resolution: Greater of ±3% of reading (not to exceed 1% of
range for % range Analyzers) or ±0.02% of range (except ranges less than or
equal to100 ppm, ±3% of reading or ± 0.05% of range).
RESPONSE TIME
Typically less than 10 seconds to read 90% of a step change. Equilibrium
time depends on the specific conditions.
OXYGEN SENSITIVITY
3 ppb (310E-H0050M Model only)
LOW DETECTION LIMIT
3 ppb (310E-H0050M Model only)
RESOLUTION
Mo
del
Number
Range
Auto
Scale A*
Display
Auto
Scale B*
Display
S00050
0
-
50 ppm
0
-
50 ppm
S00100
0
-
100 ppm
0
-
100 ppm
S00500
0
-
500 ppm
0
-
500 ppm
S01000
0
-
1000 ppm
0
-
1000 ppm
S05000
0
-
5000 ppm
0
-
5000 ppm
S10000
0
-
100
00 ppm
0
-
10000 ppm
XXXXX
S000P5
0
-
5 %
0
-
5 %
X.XX
S00P10
0
-
10 %
0
-
10 %
S00P25
0
-
25 %
0
-
25 %
XX.X
H0050M
0
-
50 ppm
0
-
5
5
-
50
H00100
0
-
100 ppm
0
-
10
10
-
100
H00500
0
-
500 ppm
0
-
50
XX.X
50
-
500
XXX.
H01000
0
-
1
000 ppm
0
-
100
100
-
1000
H05000
0
-
5000 ppm
0
-
500
XXX.
500
-
5000
H10000
0
-
10000 ppm
0
-
1000
1000
-
10000
XXXXX
H000P5
0
-
5 %
0
-
0.5
.XXX%
0.5
-
5
X.XX
H00P10
0
-
10 %
0
-
1
X.XXX%
1
-
10
H00P25
0
-
25%
0
-
2.5
X.XX%
2.5
-
25
XX.X
*Scale A applies to High Resolution models only. Scale B extends down to 0
ppm or 0% on Standard Resolution models.