Specifications DF-310E 9
3 Specifications
PERFORMANCE
ACCURACY
Standard Resolution: Greater of ± 3% of reading (not to exceed 1% of
range for % Analyzers) or 0.5% of range.
High Resolution: Greater of ±3% of reading (not to exceed 1% of
range for % range Analyzers) or ±0.02% of range (except ranges less than or
equal to100 ppm, ±3% of reading or ± 0.05% of range).
RESPONSE TIME
Typically less than 10 seconds to read 90% of a step change. Equilibrium
time depends on the specific conditions.
OXYGEN SENSITIVITY
3 ppb (310E-H0050M Model only)
LOW DETECTION LIMIT
3 ppb (310E-H0050M Model only)
RESOLUTION
Number
Scale A*
Scale B*
*Scale A applies to High Resolution models only. Scale B extends down to 0
ppm or 0% on Standard Resolution models.