Specifications DF-310E 11
Meets CISPR – 11(90) Class B Group 1 Standard
CONSTRUCTION
WEIGHT
9.5 lbs. (4.3kg) Standard Model (no options)
DIMENSIONS - Overall
8.375"w x 8.0"h x 8.5"d (21.3 cm x 20.3 cm x 21.6 cm) (with handle and gas
fittings)
CE Approved
CSA Approved
GAS SAMPLE CONDITIONS
GAS CONNECTIONS
1/8" Compression inlet and outlet Standard
1/4" Metal-face-seal inlet (Optional)
SAMPLE INLET PRESSURE
0.2 psig to 1.0 psig; 5 - 15 psig with VCR welded sample inlet (orifice
restricted)
SAMPLE FLOW RATE
1.0 to 2.0 scfh standard operating limits
GAS COMPATIBILITY
Standard Sensor: All inert and passive gases, including N
2
, H
2
, CO, Ar,
freons, hydrocarbons, etc.
STAB-EL Sensor: Limited tolerance to gas compositions containing "acid"
gases such as CO
2
, H
2
S, Cl
2
, NO
x
, SO
2
, HCl, etc.
GAS SAMPLE MOISTURE CONTENT
No limit (avoid condensation)
OIL/SOLVENT MIST
<0.03 mg/L Standard limit
>0.03 mg/L Use filter
SOLID PARTICLES
<0.01 mg/L Standard limit, Use filter if >0.01 mg/L

Table of Contents