User Interface DF-310E 81
When the composition of the gas (or the GSF factor) has been entered move the asterisk to
Update & Quit and press . The GSF will be calculated and displayed.
If the GSF factor of the gas used to calibrate the system is already known, it can be entered
directly. To enter the GSF directly, move the asterisk to the GSF line and press . Use the
and keys and hit to enter the desired value.
NOTE
The GSF for the gas used to calibrate the system may be different
from that used during analysis. If the GSF is changed to reflect the
composition of the calibrating gas, be sure to reset the GSF before
analyzing samples.