EasyManuals Logo

Servomex SERVOPRO FID User Manual

Servomex SERVOPRO FID
67 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 11
2 SPECIFICATION
WARNING
You must install and use the SERVOPRO FID in accordance with
the requirements of this section and subsequent sections of the
manual. If you do not, the protection facilities incorporated into the
design of the SERVOPRO FID may not operate as intended,
sample gas measurements may not be accurate, or the SERVOPRO
FID may be damaged.
2.1 General
Dimensions:
Height x width x depth:
133 x 483 x 457 mm (5.25 x 19 x 18 ins)
Weight (maximum)
12 kg (26 lbs)
Electrical supply requirements:
Voltage:
Frequency:
Maximum power consumption:
100 to 120 Vac or 220 to 240 Vac
2
,
3
50 to 60 Hz
150 VA
Fuse rating/type:
Two fuses 1 amp / 250 V, Type “S”, Slo-
Blo miniature fuses 5 x 20 mm for 220-
240 VAC version.
Two fuses 2 amp / 250 V, Type “S” Slo-
blo miniature fuses 5x20 mm for 100-120
VAC version
Gas flow rate
4
Minimum
Air:
Fuel:
Sample:
Maximum:
Air:
Fuel:
Sample:
300 ml min
-1
40 ml min
-1
30 ml min
-1
500 ml min
-1
100 ml min
-1
100 ml min
-1
Table 2 : General specifications
2
The analyser is supplied configured for operation with one of these voltage ranges. You must
specify the voltage range when you order the analyser.
3
This analyser is rated in accordance with Over Voltage Category II, Pollution Degree 2.
4
Flow rates are background dependent. They must be set as specified in section 2.4.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO FID and is the answer not in the manual?

Servomex SERVOPRO FID Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO FID
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals