SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 28
Actual sample sensor signal.
Actual sample pressure regulator
control output.
Actual FID. temperature sensor’s
signal.
The power of the FID heater.
User interface software version.
IO Board software version.
Sample pressure regulator software
version.
Air pressure regulator software
version.
Fuel pressure regulator software
version.
Date of the last maintenance.
Serial number of analyser.
Table 18 : Diagnostic parameters
The alarm historic menu is used to report alarms and system errors. Every time an
alarm or an error occurs, a message is displayed with the time and date. When the
error is resolved, the same message is displayed with time and date as well as:
OK. For example if a flame off occurs and then the error is resolved, flame off:ok
will be displayed.