SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 12
CAUTION
Always use factory flow set points. Excessive flows could damage detector.
Flow set points shall only be changed by a trained technician.
2.2 Gas
Pressure range
Air:
Fuel:
Sample:
103.4 to 206.9 kPa (15 to 30 psig)
103.4 to 206.9 kPa (15 to 30 psig)
34.5 to 103.4 kPa (5 to 15 psig)
Table 3 : Gas pressure specifications
2.2.1 Type of fuel
Refer to table 1 for combinations of Background and Fuel