SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 14
2.5 Typical gas settings for ignition
Argon, Nitrogen, Air, Helium, CO2
Air
Fuel
Sample
500 ml min
-1
70 ml min
-1
50 ml min
-1
Oxygen
Air
Fuel
Sample
250 ml min
-1
100 ml min
-1
10 ml min
-1
Table 6 : Ignitions settings
2.6 Environmental limits
Ambient temperature range
+5°C to +40°C
Operating ambient humidity range
0 to 95% RH non-condensing
Maximum operating altitude
2000 m (max)
Ingress protection
IP20
Table 7 : Environmental limits