SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 13
2.3 Calibration gases
Zero calibration gas
5N Hydrocarbon free sample
5,6
Span calibration gas
Sample gas containing methane
7,8
Calibration gas flow rate
Minimum
Maximum
10 ml min
-1
150 ml min
-1
Calibration gas pressure range
Minimum
Maximum
34.5 kPa (5 psig)
103.4 kPa (15 psig)
Table 4 : Calibration gas specifications
2.4 Gas settings during normal operation
Air
Fuel
Sample
300 to 500 ml min
-1
40 to 100 ml min
-1
30 to 100 ml min
-1
Table 5 : Gas settings during normal operation
5
The zero calibration gas must be of the same type as the ordered background. For example, a
SERVOPRO FID measuring hydrocarbons in argon must be calibrated with a pure argon sample
gas.
6
A hydrocarbon trap must be used in order to remove all hydrocarbons.
7
The span gas must of the same background as the ordered background.
8
The suggested amount of methane in the sample is at least 80% of the range in use. If the range
0-1000 ppm is used, a sample containing at least 800 ppm of methane is required.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO FID and is the answer not in the manual?

Servomex SERVOPRO FID Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO FID
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals