7 FAULT FINDING
Burner doesn’t ignite
burner jet diluted with other
gases because fuel system is
Purge fuel gas regulator.
Air and fuel flow not
properly set.
Adjust air and flow set
points.
Analyser not able to reach
starting flow set points.
Verify all inlet pressures.
Analyser’s reading is noisy
Air or fuel are
contaminated.
Replace air or fuel cylinder.
Add hydrocarbon traps.
There is a leak on the
sample line.
Leak test your sampling
system.
Flows are not stable.
Verify that all gas pressures
are stable.
Verify that flow readings
are stable.
Optimize PID loop
gases to the SERVOPRO
FID were not properly
Clean tubing.
There is a dead volume on
one of the gas stream.
and remove any dead
Exhaust is blocked by
water.
Make sure the exhaust
tubing has a slope.
Loss of sensitivity or no
signal
burner partially short
circuited by combustion
Clean burner.
Polarisation spring is not in
contact with FID jet.
verify that the spring is in
False flame indication
Flame sensor may be
damaged.
Change flame sensor.
Table 20 :Fault finding