SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 62
APPENDIX
A 6 FUEL SHUT-OFF/Hydrogen relay output
WARNING
A fuel shut-off valve must be connected to the SERVOPRO FID to
prevent the accumulation of fuel. The inside control valve is not
considered tight shut-off.
A fuel shut-off valve can be connected to the rear connector of the analyser. This
valve must be normally close. This valve is controlled by the SERVOPRO FID
software. When an error occurs, it closes this valve automatically to prevent the
accumulation of fuel inside the SERVOPRO FID. See section 9.1 to know more.