6.3.8 Protection fuses
Protection fuses are foreseen in the following circuit:
• Main power supply
• Supplementary outlet (max. 6 A)
• 12Vdc led power supply
• 5Vdc power supply for valves group heating
• 5Vdc power supply for elettronics
• 12Vdc power supply for flowmeter box
• 12Vdc ventilator power supply
• 12Vdc battery power supply
WARNING! Risk of injury for the user / patient
The operations described here below must be carried out by highly
qualified technical personnel, trained and formally authorised by SIARE
only.
In case of break of protection fuse:
• interrupt the mai power supply;
• eliminate the abnormal or the cause which caused the
fuse break,
• replace the protection fuse with one with the same value
and technical characteristics.
WARNING! Risk of injury for the user / patient
Fuses with wrong values and technical characteristics can compromise
equipment integrity and safety.