Preparation to use
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
6-35
6.5.1 Preliminary checks
O
2
CELL CALIBRATION
WARNING! - Risk of anaesthesia unit failure and/or injuries for
the patient.
This test verifies the correct O
2
cell operation
WARNING! - Risk of injury to the patient.
Perform the O
2
CELL CALIBRATION : consult the “Maintenance“
chapter.
• STAND-BY operative mode.
• Oxygen cell correctly positioned and
connected.
• Press for a few seconds the CAL membrane
key.
• The oxygen cell calibration procedure
begins.
• The message CAL indicates that the oxygen
cell calibration phase is in progress.
• The remaining time at the end of oxygen cell
calibration appears on PAW display.