Preparation to use
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
6-35
6.5.1 Preliminary checks
O
2
CELL CALIBRATION
WARNING! - Risk of anaesthesia unit failure and/or injuries for
the patient.
This test verifies the correct O
2
cell operation
WARNING! - Risk of injury to the patient.
Perform the O
2
CELL CALIBRATION : consult the “Maintenance“
chapter.
STAND-BY operative mode.
Oxygen cell correctly positioned and
connected.
Press for a few seconds the CAL membrane
key.
The oxygen cell calibration procedure
begins.
The message CAL indicates that the oxygen
cell calibration phase is in progress.
The remaining time at the end of oxygen cell
calibration appears on PAW display.