Preparation to use
6-36 MORPHEUS MRI
A FiO
2
value equal to 99% is displayed at
the end of calibration procedure.
At the end of calibration procedure, if all has been correctly
performed, a percentage value of FiO
2
equal to 99% shoould be
displayed.
For a correct calibration it is necessary that the cell is mounted in the
apposite location and electrically connected by the apposite cable.
If one of these conditions have not been effected the calibration is
not correct.
REPLACEMENT OF OXYGEN CELL
The oxygen cell must be replaced when at the end of calibration
phase the measured value is indicatively lower than 20% and/or in
case of displayed alarm message
For ordering the spare cell and for disposal of depleted one consult
the “ Maintenance“ chapter.