AD.A module
5-2 MORPHEUS MRI
5.1 Description
START-UP PROCEDURE OF THE VENTILATOR
• Place the main power switch in position I (ON) ; anaesthesia
unit power on.
• On the ventilator front panel, verify the electric power supply
led; green led on.
• Keep STAND-BY / ON-OFF button pressed for a few seconds
to turn the ventilator module on.
The MORPHEUS anaesthesia unit “ SELF TEST “ phase starts.