Siare Morpheus Series - Controls for Preliminary Checks

Default Icon
221 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AD.A module
5-8 MORPHEUS MRI
5.2.2 Controls for Preliminary checks
O
2
CELL CALIBRATION
WARNING! - Risk of anaesthesia unit failure and/or injuries for the
patient.
This test verifies the correct O
2
cell operation
WARNING! - Risk of injury to the patient.
Perform the O
2
CELL CALIBRATION : consult the “Maintenance“
chapter.
STAND-BY operative mode.
Oxygen cell is positioned and correctly
connected.
Press for a few seconds the CAL membrane
key.
The oxygen cell calibration procedure starts
(about 45 sec.).
The message CAL indicates that the
calibration phase of oxygen cell is in
progress.
On PAW display appears the remaining time
at the end of oxygen cell calibration.