AD.A module
5-8 MORPHEUS MRI
5.2.2 Controls for Preliminary checks
O
2
CELL CALIBRATION
WARNING! - Risk of anaesthesia unit failure and/or injuries for the
patient.
This test verifies the correct O
cell operation
WARNING! - Risk of injury to the patient.
Perform the O
2
CELL CALIBRATION : consult the “Maintenance“
chapter.
• STAND-BY operative mode.
• Oxygen cell is positioned and correctly
connected.
• Press for a few seconds the CAL membrane
key.
• The oxygen cell calibration procedure starts
•
The message CAL indicates that the
calibration phase of oxygen cell is in
progress.
• On PAW display appears the remaining time
at the end of oxygen cell calibration.