EasyManua.ls Logo

Siare Morpheus Series - 9 Troubleshooting

Default Icon
221 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AD.A module
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
5-11
The losses value of anaesthesia unit is displayed in ml basing on known
and calculated data (flow, pressure and time),
The displayed value must be less than 100 ml.
The values obtained during the LEAK TEST are displayed for the
“compensation of dead space” function.
Leakage TEST not overcome
WARNING Risk of failure of anaesthesia unit.
This test is not overcome when the pneumatic circuit pressure does not
reach 30 cmH
2
O ( code
E1 ) or when there is a leakage over 100 ml (
code
E2 ).
If the preliminary checks phase is not overcome: see cfr. 8 (Alarms)
and cfr. 9 (Troubleshooting).

Table of Contents

Related product manuals