Introduction
User’s Manual, DU3310103-MRI / rev. 3 - 02.05.2019 1-3
The CO
2
absorber canister is located in the upper side and it easy to disconnect
by apposite unlock lever. With canister inserted the system makes automatic
configuration in rebreathing modality; with taken off canister, the system makes
automatic configuration in non rebreathing modality (real open circuit). It is
possible to put and take off the canister during intervention. The canister is
available in version for reuse and disposable version including soda lime
granules.
Low periodic and extraordinary maintenance operations; easy training of
technical personnel thanks to the extreme rationality of the system and to the
drastic reduction of accidental leaks.
1.3.3 Flowmeter box
The flowmeter box is available at five rotameters; it is characterized by the complete
absence of mechanical controls; replaced by a small control panel located on the left
side of the work-shelf to manage all manual ventilation operations. This solution
allows the presence of audible and visual alarms of strong impact and without
duration limits.
1.3.4 Lung ventilator module - DISPLAY
The lung ventilator module, through user’s interface, allows to set and display the
main respiratory parameters useful for the operator to use the MORPHEUS LT/MRI
anaesthesia unit.
The lung ventilator module foresees the following operative modes: MANUAL, VC-
VAC, PCV-APCV. This ventilator type is characterized by a very simple user’s
interface and by an important novelty relevant to the presence of PRESET programs
for adults, children and newborns.
The passage from MANUAL to AUTOMATIC ventilation is made by simple pressure
on the push-button on the keyboard alongside the DISPLAY.
The valves group is automatically configured for the selected modality without any
manual procedures, avoiding errors or inadequate manoeuvres.
The lung ventilator module, foresees the possibility to ad just the end expiration
positive pressure (PEEP), the trigger sensitivity and is equipped with the FiO
2
monitoring with automatic calibration and leak test.