Preparation to use
6-6 MORPHEUS MRI
6.2.2 Flow sensor assembling
Unpack carefully the flow sensor and
the connecting hoses for airways
pressure detection.
Insert the flow sensor in the EXP
connector of the valves group.
Fix correctly the hoses for pressure
detection from sensor to the
connectors on the anaesthesia unit.
Position the hoses for detection of
airways pressure on the patient circuit
supporting arm.
The flow sensor is single patient type at differential pressure.
Consult the chapter 5 for flow sensor calibration instructions.
WARNING: Danger of injury for the patient.
Check the flow sensor connection
.
White hose connected to valves group side
Blue hose connected to patient group side