Alarms: AD.A module
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
8-9
8.4.1 Regulations and default values table
The available regulations and the default regulations which can be
selected by the operator are listed here below.
Gas supply
No setting
High PAW
This entry allows to set the high pressure alarm limit:
• 10 - 80 cmH
2
O (VAC operative mode)
Low PAW
Pre-set value
• 5 cmH
2
O
FiO2%
Pre-set value
• 19%
Apnoea
Pre-set value
• It activated with RATE = 0
Vte
Settable value
• LOFF, L 10% ……… L 90% (step 10% ; by soft key Vt)
• Default , L 50%
Power failure
(main power
supply)
Pre-set values
100% , 50% , 10%
WARNING! Risk of injury for the user / patient
The anaesthesia units used in the same health environments can
have different preset configurations of alarm limits.
• Verify that the preset alarm limits are suitable for the new
patient.