Preparation to use
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
6-39
6.5.2 Preliminary checks Flowmeter box
The preliminary tests have the aim to verify the correct connection and functionality of
anaesthesia unit and of all its modules.
Before using the anaesthesia unit on a patient it is necessary to
perform a series of preliminary tests to verify the correct unit
operation.
The list of preliminary tests is available at the end of the present
chapter.
The preliminary tests must be performed:
every time that the anaesthesia unit is turned on and used
every time that is effected a connection or an important
component is replaced (e.g., valves group, patient circuit, oxygen
cells)
Supply the anaesthesia unit through the main switch (cfr. 6.3.9).
Turn the ventilator on by the membrane key on front side of
ventilator (cfr. 6.3.9).
STAND-BY operative mode.