Preparation to use
6-38 MORPHEUS MRI
The losses value of anaesthesia unit is displayed in ml basing on
known and calculated data (flow, pressure and time),
The displayed value must be less than 100 ml.
The values obtained during the LEAK TEST are displayed for the
“compensation of dead space” function.
Leakage TEST not overcome
WARNING – Risk of failure of anaesthesia unit.
This test is not overcome when the pneumatic circuit pressure does
not reach 30 cmH
2
O ( code
E1 ) or when there is a leakage over
100 ml ( code
E2 ).
If the preliminary checks phase is not overcome: see cfr. 8 (Alarms)
and cfr. 9 (Troubleshooting).