Preparation to use
6-34 MORPHEUS MRI
6.5 Preliminary checks – Operating phase
The anaestesia unit must be ON, in STAND-BY operative mode,
once overcome the preliminary tests.
The preliminary tests aer divided in four categories:
1. TEST ON DEMAND
• O
2
CELL CALIBRATION
•
LEAK TEST
2. Flowmeter module
3. Ventilator module
•
Spirometry
• Respiratory parameters
4. Alarms
WARNING! - Risk of anaesthesia unit failure and/or injuries for
the patient.
A performance or cancellation of preliminary tests could determine a
malfunctioning during the unit operation: pay much warning.
Unless in case of emergency situation, always follow all preliminary
tests.
If the preliminary tests are not performed due to an emergency
situation, they could be performed in the more complete way as soon
as possible.