Valves group
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
3-7
3.4 Use
When the CO
2
absorber canister is inserted, the system is
automatically configured in rebreathing modality.
When the CO
2
absorber canister is removed, the system is
automatically configured in non-rebreathing modality.
3.4.1 CO
2
soda lime absorber canister
The CO
2
absorber canister is positioned in the upper side of valves group; it can
be easily disconnected by apposite lock lever, this functionality make it
replaceable also during surgical interventions.
When the CO
2
absorber canister is inserted, the system is automatically
configured in rebreathing modality.
When the canister is removed, the system is automatically configured in non
rebreathing modality (real open circuit).
It is possible to insert and remove the canister during interventions. The canister
is available in reusable or pre-loaded disposable versions.