Maintenance
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
10-5
10.2.2 Disinfection and sterilisation
To disinfect the components, dismantle them and place them in a steam disinfection
chamber at 93°C for 10 minutes.
After this first operation and before placing the components in an autoclave, wrap
them in muslin or in a similar material.
Effective sterilization is achieved in an autoclave at 121°C for approx. 15 minutes.
WARNING! Risk of injury for the patient
Always refer to the instructions provided by the autoclave
manufacturer regarding temperature and time.
• Do not disinfect, sterilize or re-use disposable products.
• Disinfect and sterilize every time an infected patient is ventilated.
• In normal conditions, disinfect and sterilize according to how often the machine is
used and in any case at least once a month.
Siare recommends:
• that the components should be checked every time they are
sterilized and any damaged parts should be replaced.
• carrying out a functioning test of the machine whenever parts or
components are replaced.