7.2.2 Administration of fresh gas in the "TO and FRO" system
In order to deliver the fresh gas flow to the TO and FRO system it is necessary that:
the lung ventilator is in MANUAL operative mode;
the TO and FRO patient circuit is connected to the AUX
connector;
that the fresh gas exit selection is activated : enabling of
fresh gas exit connector (AUX) on control panel for
manual ventilation;
that the flowmeter box is correctly set.
CAUTION. Risk of injury to the patient.
In this operative modality there is no monitoring of respiratory vital
parameters: all the patient ventilation phase is entrusted to the User’s
sensitivity.
The fresh gas flow must be high to avoid CO
2
accumulation.