Preparation to use
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
6-5
6.2 Before the use
6.2.1 Assembling of O
2
cell
WARNING. Danger of injuries to the user
To avoid the danger of electric shocks and/or break of component
parts, during interventions, ensure that the electric power supply
disconnected from anaesthesia unit.
Unpack carefully the O
2
cell
Lift the valves group cover.
Insert and screw the cell, in the space
provided with the “O2 SENSOR” script.
Ensure that the electric cable for cell
connection is positioned in the
dedicated space.
Connect the plug on the O
2
sensor.
Connect the RJ connector on the
dedicated intake within the valves
group.
WARNING! Risk of injury for the patient
Check of electric connection of O
2
cell.
At each anaesthesia unit start-up, the system check the presence of
the electric connection to the O
2
cell.
“SELF TEST” phase ” (cfr. 6.4.1) or cfr. 7.