Appendix
11-14 MORPHEUS MRI
11.4 Electromagnetic compatibility tables
11.4.1 Annex A: Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The Morpheus LT is intended for use in the electromagnetic environme
nt specified below. The customer or
the user of the Morpheus LT should assure that it is used in such an environment.
Emissions test Compliance
Virdict
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 Compliance
The Morpheus LT uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A Compliance
The Morpheus LT is suitable for use in all
establishments other than domestic, and
may be used in domestic establishments
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes, provided the following warning
is heeded:
Warning: This equipment/system is
intended for use by healthcare
professionals only. This equipment/
system may cause radio interference or
may disrupt the operation of nearby
equipment. It may be necessary to take
mitigation measures, such as re-orienting
or relocating the Morpheus LT or
shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Class A, B, C, D, o
NOT APPLICABLE
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance Compliance