Preparation to use
6-40 MORPHEUS MRI
Flowmeter box S5
Manometers
Verify that the pressures of medical gases
displayed in the manometers are from 280 kPa to
600 kPa (2,8 - 6 bar).
MIX-LIFE (device used to avoid the
administration of hypoxic mixtures).
Activate the enabling control (9.4) of valves
group.
Enable and open the N
2
O (4.7) regulator at 6
l/min on the flowmeter (4.2).
Verify that with O
2
reculator closet you read
on flowmeter (4.1) a passage included
between 1.5 and 3 l/min.