Preparation to use
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
6-49
Power failure During the anaesthesia unit operation place on OFF (
0 ) the main switch (cfr. 6.3.9).
The power failure alarm occurs: silence the alarm.
The indication “ Power failure“ is activated.
Replace the main switch in ON position ( I ).
Gas failure During anaesthesia unit operation close the medical
gas supply
The gas supplì failure alarm occurs: silence the alarm.
Restore the medical gas supply.
Apnoea This alarm is activated when the RATE value is equal
to 0.
Activate the alarm disconnecting the patient circuit
from patient simulator.
After around 20 sec. the alarm is activated, then
silence the alarm.
Vte This alarm is activated when the tidal volume value is
higher (lower) than % value of max (minimum) limit
set by operator.
Activate the alarm creating a leak in the patient circuit
higher than set alarm limit .
After around 20 secs. the alarm is activated, the
silence the alarm.
WARNING! ! Risk of injuries for the patient
The alarms must be activated in the correct times and modes.
Verify the correct activation of the visual and acoustic indications.