AD.A module
7-4 MORPHEUS MRI
High flows open circuit ( NON REBREATHING )
To perform this type of ventilation it is necessary to take off the soda
lime canister from valves group.
The breathing gas concentration is similar to the fresh gas ones.
There is a high consumption of fresh gas
The temperature and humidity of breathing gas are low. The high flow
ventilation is recommended only for pre-anaesthesia and awakening or for brief
surgeries.
There is not CO
2
accumulation in the patient circuit.
Required supervision: normal
Semi-closed circuit ( REBREATHING )
The fresh gas flow in the patient circuit is lower than the ones breathed by the
patient; therefore the fresh gas concentration of the breathed gas is different
than the ones of the fresh gas, related to the BREATHED GAS / FRESH GAS
ratio
There is a medium consumption of soda lime.
There is a medium consumption of fresh gas.
The temperature and humidity of the fresh gas are acceptable for a long surgery
situation.
Required supervision: pay attention to the gas concentration
monitoring (O
2
, N
2
O, CO
2
and halogenated gas).