Troubleshooting
9-2 MORPHEUS MRI
Initialization
The initialization phase is not completed and the system is blocked.
• Verify and intervene in function on the error messages and indications
evidenced during the “SELF TEST” phase.
• Turn off and on the anaesthesia unit and repeat “SELF TEST” phase.
• Contact the Siare Service Centre or a Centre authorised by Siare.
Keys and
encoder knob
The keys or the encoder knob do not work
• Switch the anaesthesia unit OFF and then switch back ON.
• Contact the Siare Service Centre or a Centre authorised by Siare.
Low gas
pressure
This alarm is activated when the pressure is insufficient (< 2.7 bar) for
the anaesthesia unit to operate correctly.
• Check that the medical gases are correctly connected to the
anaesthesia workstation. Restore the connections or replace the tubes if
damaged.
• Check that there is sufficient pressure in the supply system or in the
cylinders. Adjust or repair the supply system (or replace the cylinders) if
the pressure is insufficient.
• Contact the Siare Service Centre or a Centre authorised by Siare.
Battery
charge level
This alarm is activated when the charge level of the battery is at 50%
of the fully charged level: no more than 60 minutes of operating
autonomy is guaranteed.
• Check that it is connected to the main power supply.
• Recharge the battery.
• If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention of a