Troubleshooting
9-2 MORPHEUS MRI
Initialization
phase
The initialization phase is not completed and the system is blocked.
Verify and intervene in function on the error messages and indications
evidenced during the “SELF TEST” phase.
Turn off and on the anaesthesia unit and repeat “SELF TEST” phase.
Contact the Siare Service Centre or a Centre authorised by Siare.
Keys and
encoder knob
The keys or the encoder knob do not work
Switch the anaesthesia unit OFF and then switch back ON.
Contact the Siare Service Centre or a Centre authorised by Siare.
Low gas
pressure
This alarm is activated when the pressure is insufficient (< 2.7 bar) for
the anaesthesia unit to operate correctly.
Check that the medical gases are correctly connected to the
anaesthesia workstation. Restore the connections or replace the tubes if
damaged.
Check that there is sufficient pressure in the supply system or in the
cylinders. Adjust or repair the supply system (or replace the cylinders) if
the pressure is insufficient.
Contact the Siare Service Centre or a Centre authorised by Siare.
Battery
charge level
This alarm is activated when the charge level of the battery is at 50%
of the fully charged level: no more than 60 minutes of operating
autonomy is guaranteed.
Check that it is connected to the main power supply.
Recharge the battery.
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention of a
Service Centre.