Maintenance
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
10-9
Breathing
system
Dismantle and clean, then sterilize in an
autoclave, disinfect with steam or
chemically.
Before using again,
eliminate any humidity
inside the components
by means of
compressed air.
Flow sensor
Single patient use type at differential
pressure
It is possible to sterilize
with gamma rays or with
ethylene oxide (ETO), at
first use.
WARNING. Risk of device failure.
Replace the sensor according to the procedures and local
ref. directives and in function of use of filters on patient
circuit.
Other
accessories
Carefully follow the manufacturer’s
instructions.
Refer to the
accompanying
documentation.
Electrical
connections
On the aim to guarantee patient and
operator safety it is necessary to keep the
power supply cable in perfect conditions.
Perform daily checking’s
of cable condition; any
damage, also a
minimum damage, must
be promptly eliminated,
eventually replacing the
whole cable.