Setting of Vte alarm limit
WARNING!
After having performed the “Flow Sensor Calibration” it is necessary,
each time, to set the value, in %, of Vte alarm limit.
• LOFF, from L 10% to L 90% (step 10% )
• Default , L 50%
To set the Vte alarm limits, consult the chapter 8, Alarms.
Verification of spirometry
VAC operative mode:
• Basing on set ventilation parameters check carefully the numerical
data and the parameters calculated by the flow sensor on the
expiratory line.
• Vary the set values of the main parameters and check their
correspondence on measured parameters (± 10% tolerance).