AD.A module
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
5-23
Setting of Vte alarm limit
WARNING!
After having performed the “Flow Sensor Calibration” it is necessary,
each time, to set the value, in %, of Vte alarm limit.
LOFF, from L 10% to L 90% (step 10% )
Default , L 50%
To set the Vte alarm limits, consult the chapter 8, Alarms.
Verification of spirometry
VAC operative mode:
Basing on set ventilation parameters check carefully the numerical
data and the parameters calculated by the flow sensor on the
expiratory line.
Vary the set values of the main parameters and check their
correspondence on measured parameters (± 10% tolerance).