FLOWSIC600 · Operating Instructions · 8010458 V 2.0 · © SICK MAIHAK GmbH 19
Product Description
****The “active” or “inactive” state can be assigned to the electric switch status
“normally open” or “normally closed” by configuration in the MEPAFLOW600 CBM software
(adjust settings for register #5101 on the “Parameters” page.)
The LCD display is capable of displaying measured values, parameters, messages and
other information.
A flashing letter at the upper right corner of the LCD display indicates that a logbook
contains unacknowledged logbook entries. Depending on the type of entry this will be:
“I” for Information
“W” for Warning
“E” for Error
After acknowledging all new entries, the flashing letter disappears. For details see
pg.68,
5.4.1.
Table 2 Status output
Output signal / LCD / port
Signal behavior
Measurement state Check request state Configuration Mode Measurement invalid
“Check request”
Status signal
Status
“active / inactive” ***
Measurement valid
Status
“active / inactive” ***
Compensation of path
failure
“undefined” “undefined”
“Direction of flow”
Status signal
Status
“active / inactive” ***
Positive or negative
direction of flow
Status
“active / inactive” ***
Positive or negative
direction of flow
“undefined” “undefined”
“Warning” Status
“active / inactive” ***
Status
“active / inactive” ***
“undefined” “undefined”
LCD display
Display flashing Display flashing
Serial port RS485
Measured value, diagnosis information and parameters
Measuring data logging, diagnosis and configuration through the MEPAFLOW600
CBM software
Connection with external process control equipment through implemented
MODBUS protocol (data polling)
+V 123456 m³
-V 1234 m³
1234 m³
FLOWSIC600
Configuration
+V 123456 m³
-V 1234 m³