Maintenance
64 FLOWSIC600 · Operating Instructions · 8010458 V2.0 · © SICK MAIHAK GmbH
5.2.2 Checking the Meter Health
The FLOWSIC600 monitors its own meter health with User Warnings and system alarms. If
the outputs are configured to indicate alarms and / or User Warnings, it is not necessary to
manually check the meter health.
If you want to get a visual feedback about your meter‘s health, the “Main system bar“ in
MEPAFLOW600 CBM provides a compact overview:
b
Use MEPAFLOW600 CBM to connect to your meter (
pg.47, 4.3).
b
Check the main system bar for any yellow or red icon (
Figure 27). A red or yellow icon
indicates a potential problem with the meter.
If any of the icons in the main system bar are yellow or red, proceed with checking the
“Meter Status” (
pg.73, 6.2.1) and the “User Warnings” (
pg.75, 6.2.2).
Figure 27 “Main system bar”
Main system bar
Measured SOS
Icons for system alarms,
User Warnings and
performance

Table of Contents