EasyManua.ls Logo

SICK FLOWSIC600

SICK FLOWSIC600
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
36 FLOWSIC600 · Operating Instructions · 8010458 V2.0 · © SICK MAIHAK GmbH
3.4.4 Terminal enclosure on the SPU
Opening the rear housing cover
b
Loosen the securing clip using a 3 mm Allen key.
b
Turn the rear housing cover counter-clockwise and take it off.
A schematic wiring diagram is provided on the inside of the rear housing cover.
Figure 10 SPU housing
Figure 11 Terminal box on the rear of the SPU (see Section
3.4.2 for North American wiring specification equivalents)
NOTICE: Lubricant
Only use LOCTITE 8156 as lubricant for front and rear housing cover.
Cover for power
supply terminals
Cable entry for internal 10-
core cable
10-pole terminal block
for signal inputs and outputs
1
2
HSK-K type cable glands,
M 20 x 1.5 plastic (EU)
or ½ in NPT (North America)
Power supply
2 x 1.5 mm
2
(LiYCY or equivalent)
Digital output / current
output
4 x 2 x 0.5 mm
2
(Li2YCY [TP] or equivalent)
MODBUS
4 x 2 x 0.5 mm
2
(Li2YCY [TP] or equivalent)
Bridge

Table of Contents

Other manuals for SICK FLOWSIC600

Related product manuals