24
8012704/YHS4/V2-0/2016-10 | SICKOPERATING INSTRUCTIONS | GM32
Subject to change without notice
3 PREPARING THE GAS DUCT SIDE
3.5.4 Preparing the power supply
1 Requirements to connecting cable:
Cross-section: 3 x 1.5 mm
2
Temperature class: -40 ... +85 °C
2 Provide a separate external disconnecting device for:
Connection unit (max. power input see “System: GM32”, page 65)
Purge air units ( Technical data of purge air unit)
Mark the disconnecting device as disconnecting device for the GM32.
3 Lay the electric lines of the power supply to the connection unit and connect the power
supply in the connection unit.
Always connect a protective conductor to PE.
Fig. 12: Power connection in the connection unit
Lay and connect the electrical lines to the purge air units.
Take precautions to prevent switching off the purge air supply accidentally.
Attach a clearly visible warning against accidentally switching off the separation
equipment for the purge air unit.
Power supply
100 .. 240 V / 50 .. 60 Hz
Always connect a protective
conductor to PE.
The power supply must remain switched off until the GM32 is to be put into opera-
tion.

Table of Contents

Other manuals for SICK GM32

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK GM32 and is the answer not in the manual?

SICK GM32 Specifications

General IconGeneral
BrandSICK
ModelGM32
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals