7 CLEARING MALFUNCTIONS
7.2.3 Sample gas penetrating
7.2.4 Corrosion on flanges
7.2.5 Measured value blinks
If a measured value blinks: Measured value is “uncertain” (e.g. calibration range
exceeded).
7.3 Error messages
7.3.1 Example of an error message
Fig. 41: Example of an error message
NOTE: Sample gas in the analyzer can damage the analyzer.
Error Possible causes Remedy
Sample gas penetrating the
area before the SR-unit.
Gas pressure in gas duct too
high.
Check project planning.
Purge air unit has failed or is
too weak.
Check purge air unit.
Plan redundant purge air
unit.
Strengthen the purge air
unit.
Table 15: Sample gas penetrates the enclosure
Error Possible causes Remedy
Corrosion on flanges
Unsuitable materials Check project planning.
Table 16: Corrosion on flanges