50
8012704/YHS4/V2-0/2016-10 | SICKOPERATING INSTRUCTIONS | GM32
Subject to change without notice
6 MAINTENANCE
6.4 Visual inspection
Check the housings of the SR-unit, reflector unit and connection unit for mechanical
damage.
Clean the respective housings if contaminated.
Check all cables for damage.
Pay attention to chafe marks and kinks on cable ducts.
Check flanges and screw fittings for firm seating.
6.5 Cleaning the window
Fig. 35: Window of SR-unit (reflector unit accordingly)
1 Swivel the SR-unit or reflector unit out, see “Swiveling out and removing the SR-unit”,
page 49 .
2 Cleaning the window.
Use an optical cleaning cloth for cleaning.
The cleaning cloth can be moistened with demineralized water.
Do not use cleaning agents.
3 Close the SR-unit or reflector unit again.
6.6 Checking and replacing the desiccant cartridges
Fig. 36: Desiccant cartridges
1
Swivel the SR-unit out, see “Swiveling out and removing the SR-unit”, page 49.
2 Desiccant cartridge is light blue: Desiccant cartridge is dry.
Desiccant cartridge is white: Replace the Desiccant cartridge.
3 Replacing the desiccant cartridge:
a) Unscrew the desiccant cartridge.
b) Screw in the new desiccant cartridge.
4 Close the SR-unit again.
Window
2 desiccant cartridges