OPERATING INSTRUCTIONS | LFP CUBIC 8019918 /ZJA6/ 2017-07-20 | SICK AG
Subject to change without notice
57
14TECHNICAL DATA
14 Technical data
14.1 Features
Medium Liquids
Detection type Limit, continuous
Probe length
Mono-rod probe
Cable probe
200 mm … 2,000 mm
1,000 mm, 2,000 mm, 3,000 mm, 4,000 mm
Adjustable measuring range 95 mm … 6,005 mm
Process pressure –1 bar … +10 bar
Process temperature –20 °C … +100 °C
GOST certicate
RoHS certicate
IO-Link
UL certicate
14.2 Performance
Accuracy
1)
± 5 mm
Reproducibility
1)
≤ 2 mm
Resolution
< 2 mm
Response time
2)
< 400 ms
Dielectric constant
≥ 5 for mono-rod probe/rope probe
≥ 1.8 with coaxial tube
Conductivity
No limitation
Maximum level change
3)
500 mm/s
Inactive area at end of probe
1)
10 mm
1)
With water under reference conditions, see “14.7 Measurement accuracy”.
2)
Depends on measuring mode (high-speed < 400 ms, high accuracy < 2,800 ms)
3)
Depends on configuration (MaxCol - maximum change of level)