Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 41
Subject to change without notice
5.6.6.1 Measured value box (description)
A measured value box shows the measured value as numeric value.
(Measured box default: “SCU Technical Information“ manual)
Typical Measuring screen:
Touching a measured value box activates this box.
● The activated box is highlighted light brown.
– If NN (instead of component) or a.u. (instead of unit is displayed:
- No measured value has been assigned.
– If a measured value box is shown gray:
- Measured value box not used ( “SCU Technical Information“ manual).
● The status of the analyzer to which the activated (light brown) box is assigned is shown
in the status line.
Scaling (measured value box, bar graph, line writer)
Touching an activated measured value box calls up a screen to scale the measured value
box:
701
0.00
0.00 0.00
0.00
0.00
0.00 0.00
0.00
0.00
0.00
0.00
0.00
0.00
701
17.3
126
/System Control Unit/Measuring/Measuring Screen 1
SCU
MCS100FT
Operation
25.05.10
14:01
Name (e.g.: Component) -
Unit-
Status line of the analyzer for
which the measured value box
is activated (light brown) is
shown.
Measured value box color:
- Light brown: Activated
- Light blue: Valid
- Gray: Not used
Name
Unit
Name
Unit
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
Name
Unit
NN
a.u.
NN
a.u.
NN
a.u.
Name
Unit
Measured value
Measured value field color:
- White: Measured value OK
- Yellow: Maintenance request/uncer-
tain
- Red: Failure
/System Control Unit/Measuring/Measuring Screen 1
SCU
MCS100FT
Measure
black
-2
0
100
Save
Cancel
Color Choice
Data Dimension
Scale Start
Scale End
Significant positions (-99 .. +99)
Minus sign = decimal places
Example:
-2: 123.45
-1: 1234.5
0
: 12345
1
: 12340 (trailing zeros)
2 : 12300
Character color of component or
unit.
For bar graphs:
Scale start and end.
Activate
Checkmark: Display line
(Only effective for “LineWriter”)