Scheduled Maintenance
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 91
Subject to change without notice
7.2.3.1 Procedure to replace the drying agent cartridge
The MCS100FT does not have to be switched off.
Fig. 13 Replace the drying agent cartridge
1 Loosen 2 screws of the cover (Phillips screwdriver) and remove cover.
2 Pull out the drying agent cartridge.
3 Immediately insert new drying agent cartridge.
4 Reinstall the cover (curves “at the rear”) and screw tight.
Recommendation: Keep a new drying agent cartridge in storage.