Clinitek Advantus Operator’s Guide 93
System Configuration
Tests to Report and Order
Use this procedure to select the order in which analytes and physical
parameters are reported. You can choose to not report a test.
These selections apply only to testing with the primary reagent strip. They do
not apply when testing with the secondary reagent strip. Secondary reagent
results are always reported in the default order.
1. Select Tests to report and their order to display a series of cycle keys,
labeled 1 to 12.
To... Then ...
retain the existing tests and
their order,
select
Previous Screen.
change color and clarity, 1. Cycle the last active key to select
COL or
CLA.
2. Select the next key, which is now active,
to add the other test name.
Color and clarity are reported in the last 2
positions if you make no further changes.
NOTE: If English is the selected language
and S.I. is the selected Results units, color is
automatically included as the last test. You
can also add it manually to the end of the list.
You can also choose to include visually
determined clarity as a reported result.
change the order in which tests
are reported,
1. At the first position you want to change,
use the cycle key to select a test.
Any tests not already listed display first.
Then a blank displays and all tests from
that position are erased and must be re-
entered.
As each test is selected, the next test in
the list is the first test displayed for the
following position.
2. Select a test for each of the remaining
positions
3. When finished, select
Previous Screen.
remove a test from the
reporting order,
Select the tests to report and leave a blank
description in the final position.

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes

Related product manuals