Communication
11.13 DR 10 load cell parameters
SIWAREX WT241
136 Manual, 05/2015, A5E36046748A
11.13.2 50/60 Hz toggling
To improve the suppression of faults caused by the supply network, you can specify the
network frequency for signal filtering.
11.13.3 Load cell characteristic value
The characteristic value of the load cell(s) is required to correctly interpret the output voltage
from the load cell. The exact value can be entered if the measurement log for the load cell is
available. The mean value of all characteristic values is used if there is more than one load
cell.
Example
Characteristic value = 2.018 mV/V
11.13.4 Rated load of a load cell
The rated load of a load cell is required for checking the maximum weighing range of the
scales. The rated load is entered in the specified units of weight.