Messages 
  9.3 Message list 
SIWAREX WT241 
Manual, 05/2015, A5E36046748A  83 
9.3  Message list 
The message list is an overview of all messages that the SIWAREX module can generate. A 
message can be quickly identified by the message code (number). 
9.3.1  System status message list 
 
Operating errors (code 1000 to 1999) 
sorted by code no. 
1000 operating error exists 
Group message, at least one operating error exists. 
Watchdog, error is displayed for at least 10 seconds. A seri-
ous error has occurred in the function of SIWAREX, e.g. 
program error, severe electromagnetic influence on device, 
etc. 
Contact the SIWAREX Support if the error occurs multiple 
times.  
RAM error. An error has occurred in the memory; the memory 
content is no longer correct. The module must be switched 
off. If the error occurs again, SIWAREX is defective. 
1003 Checksum incorrect parameter 
Checksum error at parameter. Critical error because the 
parameters are no longer safe. 
1004 Checksum incorrect program  
Checksum error program code. Critical error because the 
program is no longer safe. 
Error when writing/deleting,or logbook full 
AD converter error when reading in the measured value. If 
the error occurs again, make sure that the EMC recommen-
dations are observed (chapter EMC-compliant setup 
(Page 20)). 
Undervoltage at sensor cables 
Overload of scale (ca. 110%) 
Underload of scale (ca. -10%) 
1107 Legal trade display failure 
The SecureDisplay legal trade display no longer communi-
cates with the module 
9.3.2  Technology error message list 
 
Technology error (code 2000 to 4999) 
Group message, at least one technology error exists 
The set recording rate for trace function cannot be processed. Set 
a slower recording rate. 
The zeroing procedure has been canceled. Possible causes: 
violation of zeroing limits. 
The trace recording has been cancelled. Possible causes: trace 
memory full and not declared as ring memory.