SIWAREX WT241 
Manual, 05/2015, A5E36046748A  81 
9.1  Message types 
The messages in the electronic weighing system described here are divided into three types.  
System status messages 
System status messages can be generated spontaneously at any time by an unexpected 
event. They include internal and external hardware problems which can occur spontaneously 
during weighing. 
Data and operating errors 
The data and operating errors are always a response to a command due to a plausibility 
check.  
These are data errors if a plausibility error has been detected in a data packet which was 
sent to the module and receipt of the packet has been rejected by the module.  
These are operating errors if the module cannot execute the sent command in its current 
operating state. 
Technology errors 
Technology errors occur spontaneously due to the process flow of a weighing. 
Status bits, on the other hand, are not messages. The status displays describe the status of 
the scale during normal operation and can be monitored or evaluated at any time.