Scale parameters and functions of the belt scale 
  8.16 Menu 5.0 User management 
SIWAREX WT241 
Manual, 05/2015, A5E36046748A  79 
8.16  Menu 5.0 User management 
The SIWAREX WT241 provides password protection. It is activated by default. The 
password protection can be deactivated (and reactivated again) via a DIP switch at the 
weighing module SIWAREX WP241 inside the terminal. 
In case the password protection is activated, menu 5.0 “User management” will be available. 
The menus 1.0 “Setup”, 3.0 “Language, Date & Time” and 5.0 “User management” are 
protected by the password. 
The default login values are: 
User:                ADMIN 
Password:  WT241 
Activate and deactivate the passwort protection by DIP switch 
Two DIP switches are located at the bottom side of the weighing module SIWAREX WP241 
just next to the Ethernet port (accessible via the air inlet).  
 
Figure 8-4 DIP switches at SIWAREX WP241weighing module 
Password protection deactivated 
Password protection activated 
As default, both switches are located in the lower position, so that the password protection is 
activated and the weighing module is in stand-alone operation mode. 
For deactivating the password protection change the position of the left DIP switch ① to top. 
It is important to keep the position of the right switch ② on bottom!  
 
Take care to keep the login data (username, password) safe. Without password there is 
not possibility to change settings or recalibrate the scale etc. In case no login data is 
available anymore, you need to reinstall the user interface with an external software.