Messages 
  9.3 Message list 
SIWAREX WT241 
Manual, 05/2015, A5E36046748A  85 
Data and operating errors 
 (code 5000 to 8999) 
7004 Assignment of DIs/DQs incorrect 
An error occurred while assigning the digital inputs or digital 
outputs. 
7007 The calibration weights or calibration 
digits are not plausible 
Specifications for calibration weights or digits in data record 
DR 3 are incorrect (minimum distance, reversal of incline). 
7008 Zeroing parameters are not plausible 
The specifications for zeroing (data record DR 3) are not 
plausible. 
7010 Scale interval / rounding 
Scale interval or selection for rounding to decimal places is 
not plausible. 
Specification of filter parameters is not plausible. 
7013 Interface assignment for calibratable 
HMI not plausible 
The assignment of the interface to the calibratable HMI is 
incorrect. 
7014 Specified time not plausible 
Specified time value is not plausible or may be signaled in 
connection with additional errors 
7016 Parameter assignment of analog 
output not plausible 
The parameters for the analog output (data record DR 7) are 
not plausible. 
7017 MAC address cannot be changed 
The specified IP addresses (DR12) are not plausible. 
7019 RS485 parameter error 
The specified RS485 interface parameters (DR13) are not 
plausible. 
7020 Speed parameter or sensor parame-
ter not plausible 
The speed parameter or sensor parameter in DR3 is not 
plausible. 
7021 Units are not plausible 
The selected units are not plausible (mixing of metric and 
imperial parameters).