6 T e chnical Description
6 - 18 [ 1 ] I N S T R U C T I O N S F O R U S E
Transducer: C6-3 3D, C6F3
NTSC Systems, Medium Sweep Speed
Measured Length
Depth
20 mm 40 mm 60 mm
80 mm within ± 11.2 % within ± 5.6 % within ± 5 %
160 mm within ± 20.0 % within ± 10.0 % within ± 6.7 %
240 mm within ± 34.5 % within ± 17.3 % within ± 11.5 %
Transducer: C6-3 3D, C6F3
PAL Systems, Standard Sweep Speed
Measured Length
Depth
20 mm 40 mm 60 mm
80 mm within ± 5.8 % within ± 5 % within ± 5 %
160 mm within ± 10.5 % within ± 5.2 % within ± 5 %
240 mm within ± 21.1 % within ± 10.6 % within ± 7 %
Transducer: C6-3 3D, C6F3
PAL Systems, Medium Sweep Speed
Measured Length
Depth
20 mm 40 mm 60 mm
80 mm within ± 11.2 % within ± 5.6 % within ± 5 %
160 mm within ± 20.0 % within ± 10.0 % within ± 6.7 %
240 mm within ± 41.3 % within ± 20.7 % within ± 13.8 %
Transducer: Endo-V 3D, EV8F5
NTSC/PAL Systems, Standard Sweep Speed)
Measured Length
Depth
20 mm 40 mm 60 mm
80 mm within ± 8.4 % within ± 5 % within ± 5 %
160 mm within ± 27.8 % within ± 13.9 % within ± 9.3 %
Transducer: Endo-V 3D, EV8F5
NTSC/PAL Systems, Medium Sweep Speed)
Measured Length
Depth
20 mm 40 mm 60 mm
80 mm within ± 11.8 % within ± 5.9 % within ± 5 %
160 mm within ± 32.1 % within ± 16.1 % within ± 10.7 %